| The system has a minimum chamber capacity of 150 litres. |
| The inner chamber and three shelves made of Type 304 stainless steel, electropolished for easy cleaning and corrosion resistance. |
| The temperature control range from 3°C above ambient to 55°C. |
| The system provide temperature setting precision of ±0.1°C and spatial uniformity within ±0.5°C. |
| CO₂ concentration adjustable from 0–20%, with ±0.1% accuracy using a thermal conductivity detector (TCD) sensor. |
| The systes has a minimum 3-litre pan-less water reservoir for humidity control. |
| An automatic sterilisation or decontamination mechanism built in. |
| The control panel microprocessor-based with a digital display for temperature, CO₂ concentration, and sterilisation cycle regulation. |
| The system has alarms for low CO₂ (drop of 0.5% below set point), low temperature (drop of 1% below set point), and power failure. |
| The system operate on 220–240 V, 50 Hz power supply. |
| We have a service facility located within 100 km of the installation site and provide valid service centre details, including contact information, with the bid submission. |
| Warranty: 03 years from the date of successful installation |